Annals of Emergency Medicine
Volume 45, Issue 6 , Pages 587-591, June 2005

Point-of-Care Testing Reduces Length of Stay in Emergency Department Chest Pain Patients

From the Department of Emergency Medicine, Stony Brook University, Stony Brook, NY

Received 27 August 2004; received in revised form 1 November 2004 and 18 November 2004; accepted 23 November 2004. published online 22 March 2005.

Study objective

We determine the effect of cardiac troponin I point-of-care testing on emergency department (ED) length of stay in chest pain patients.

Methods

This was a before-and-after trial in a university-based ED with 75,000 annual visits. Participants were consecutive patients with a chief complaint of chest pain who were admitted to the hospital. During the first 2-week period (before), only central laboratory testing of troponin was performed. During the second 2-week period (after), treating nurses performed bedside point-of-care testing for troponin I, as well as central laboratory testing. Test turnaround times, time from triage until calling in admissions, and time from triage until patients left the ED to be transferred to a floor (ED length of stay) were determined and compared between the 2 study periods. Comparisons between study periods are expressed as mean differences with 95% confidence intervals (CIs). A sample of 100 patients in each group had 90% power to detect a 1-hour difference in length of stay (2-tailed α=0.05).

Results

There were 232 patients before and 134 after introduction of point-of-care testing. Mean age (SD) was 63 years (16 years), and 44% were female patients. Baseline characteristics were similar in both groups. The rate of positive troponins was also similar (9.5% versus 6.1%). ED length of stay was significantly reduced after introduction of point-of-care testing (5.2 hours [95% CI 4.6 to 5.8 hours] versus 7.1 hours [95% CI 6.6 to 7.7 hours]; mean difference 1.9 hours [95% CI 1.1 to 2.7 hours]). The time until the admission was called in to bed control was also significantly reduced by introducing point-of-care testing (2.7 hours [95% CI 2.4 to 3.1 hours] versus 4.7 hours [95% CI 4.3 to 5.0 hours]; mean difference 1.9 hours [95% CI 1.4 to 2.5 hours]). Point-of-care testing turnaround (14.8 minutes [95% CI 14.1 to 15.5 minutes]) was significantly shorter than for central laboratory testing (83 minutes [95% CI 77 to 89 minutes]; mean difference 68 minutes [95% CI 62 minutes to 74 minutes]). With central testing as the criterion standard, point-of-care testing had a sensitivity of 100% (95% CI 63% to 100%) and a specificity of 96% (95% CI 92% to 99%).

Conclusion

Bedside performance of troponin I point-of-care testing by treating nurses significantly reduces ED length of stay.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 Author contributions: AJS was responsible for the study inception. AJS and JA conducted data analysis and manuscript drafting. AJS, JA, JG, and JC conducted study design, data collection, and manuscript review. AJS takes responsibility for the paper as a whole.Funding and support: The authors report this study did not receive any outside funding or support.Presented at the Society for Academic Emergency Medicine annual meeting, May 2004, Orlando, FL.Reprints not available from the authors.

PII: S0196-0644(04)01712-3

doi:10.1016/j.annemergmed.2004.11.020

Annals of Emergency Medicine
Volume 45, Issue 6 , Pages 587-591, June 2005