Usability Study of Two Common Defibrillators Reveals Hazards

Presented in part at the Human Factors and Ergonomics Society 48th annual meeting, September 2004, New Orleans, LA; and the National Association of EMS Physicians meeting, January 2006, Tucson, AZ.

  • Rollin J. Fairbanks, MD, MS

      Affiliations

    • Department of Emergency Medicine, University of Rochester School of Medicine, Rochester, NY
    • Corresponding Author InformationAddress for correspondence: Rollin J. (Terry) Fairbanks, MD, MS, NREMTP, Department of Emergency Medicine, University of Rochester Medical Center, 601 Elmwood Avenue, Box 655, Rochester, NY 14642; 585-273-3849, fax 585-275-2092
  • ,
  • Stanley H. Caplan, MS

      Affiliations

    • Usability Associates, Rochester, NY
  • ,
  • Paul A. Bishop, MPA, EMT-P

      Affiliations

    • Monroe Community College, Rochester, NY
  • ,
  • Aaron M. Marks, BA, EMT-P

      Affiliations

    • Monroe Ambulance, Rochester, NY.
    • Mr. Marks is now with Bridgeborn Government Solutions, Arlington, VA.
  • ,
  • Manish N. Shah, MD, MPH

      Affiliations

    • Department of Emergency Medicine, University of Rochester School of Medicine, Rochester, NY
    • Department of Community and Preventative Medicine, University of Rochester School of Medicine, Rochester, NY

Received 2 June 2006; received in revised form 3 January 2007 and 22 March 2007; accepted 27 March 2007. published online 11 May 2007.

Study objective

This usability study evaluates the user interface of 2 common monitor-defibrillators, the Lifepak10 and Lifepak12, to identify use-related hazards.

Methods

Fourteen paramedics familiar with both devices completed 4 EMS simulator scenarios using each device. The scenarios involved “quick look” and monitoring, defibrillation, synchronized cardioversion, and replacing paper. Qualitative and quantitative data were collected, including both participant self-evaluation (scored 1 to 9) and expert observer evaluation (scored 0 to 4).

Results

Participant ratings demonstrated that for performing a quick look, the Lifepak10 was easier to use (mean 8.0 versus 7.1), and for synchronized cardioversion the Lifepak12 was easier (mean 6.7 versus 5.3). Participants performed better on the Lifepak12 than the Lifepak10 for synchronized cardioversion (mean 3.1 versus 1.6) and replacing paper (mean 3.0 versus 2.1). One participant did not complete the final questionnaire. Of the remaining 13, 11 (85%) participants preferred the Lifepak12 for use on a regular basis. Eight (62%) paramedics thought that the Lifepak12 would be more effective in an emergency; 9 (69%) believed that the Lifepak10 is quicker to learn. Paramedics reported difficulty using the devices with gloves and confusion in “sync” mode. Of note, 50% of participants inadvertently delivered an unsynchronized countershock for supraventricular tachycardia.

Conclusion

Although the Lifepak10 is easier to learn, the Lifepak12 is perceived as easier to use and more effective in emergencies. The high failure rate in synchronized cardioversion indicates a need to reevaluate the user interface design for this function. Limitations of this study include the use of simulation.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 Supervising editor: Robert L. Wears, MD, MS

 Author contributions: RJF and SHC conceived the study, designed the study, and obtained research funding. RJF, SHC, and MNS supervised the conduct of the trial and data collection. RJF, PAB, AMM, and MNS undertook recruitment of participating paramedics. RJF, SHC, and MNS managed the data. All investigators participated in the qualitative analysis. MNS provided statistical advice and RJF, SHC, and MNS analyzed the data. RJF and SHC drafted the article, and all authors contributed substantially to its revision. RJF takes responsibility for the paper as a whole.

 Funding and support: By Annals policy, all authors are required to disclose any and all commercial, financial, and other relationships in any way related to the subject of this article, that may create any potential conflict of interest. See the Manuscript Submission Agreement in this issue for examples of specific conflicts covered by this statement. This research was supported by the University of Rochester Department of Emergency Medicine research seed funding. At the time of the study, Drs. Fairbanks and Shah were supported by funding from the National Institutes of Health (1R41NR009592) and Agency for Healthcare Research and Quality (1U18HS015818), and Dr. Shah, by funding from Health Resources and Services Administration.

 Available online May 11, 2007.

 Reprints not available from the authors.

PII: S0196-0644(07)00390-3

doi:10.1016/j.annemergmed.2007.03.029

Refers to article:

  • Studying the Technical Work of Emergency Care

    Christopher P. Nemeth, Richard I. Cook, Robert L. Wears
    Annals of Emergency Medicine October 2007 (Vol. 50, Issue 4, Pages 384-386)

  • When Is a Defibrillator Not a Defibrillator? When It’s Like a Clock Radio … . The Challenge of Usability and Patient Safety in the Real World , 28 August 2007

    Ben-Tzion Karsh, Matthew Scanlon
    Annals of Emergency Medicine October 2007 (Vol. 50, Issue 4, Pages 433-435)

Annals of Emergency Medicine
Volume 50, Issue 4 , Pages 424-432, October 2007